Hemostasis Analyzer for clotting and chromogenic methods Cat.No. 80 515 00
|
The TEChrom IV plus offers four different calculation
models:
The TEChrom IV Plus handle 30 different tests separate and individual with all specific calibration data. The TEChrom IV Plus fits perfectly into small and medium size Hemostasis laboratories and works excellent as a Stand-by unit to already installed automated instruments. For the usage in Emergency laboratories it is fully suitable. Tests PT, APTT, TT, FIB (Clauss/derived), Factor II, V, VII, VIII, IX, X, XI, XII, AT-III, Plasminogen, Heparin, C1-Esterase, Alpha 2-Antiplasmin, Protein C, Protein S, PAI, tPA, µPA, Heptest, Lupus Anticoagulant, Factor XIII at 340 nm, APCR D-Dimer and PT-B (whole blood PT) at 710nm, Research test - free selectable parameters, STAT with 2xPT and 2xAPTT Selectable dimensions: s, %, Ratio, INR, Units (mg/dl), dE (Slope/min), E (Extinction) etc. Technical Description The TEChrom IV Plus is a micro-computer controlled measuring device for Hemostasis tests with a 4-channel optic and a LC-Display. An additional 4-times stopwatch realizes time correct operating. The tests will be selected through keys on a foil keyboard. The dialog language is user friendly and selectable in English and German. The Main program offers following menus
An additional menu for SYSTEM ANALYSIS includes a variation of service procedures, which increases the system reliability of the instrument and also from each measurement. Any errors in the system will be detected, displayed and printed out. With the usage of the TECO Softwaretools the measuring curve data can be transmitted to a Personal Computer and loaded by any spread sheet with graphic function. Software updates are realized with an easy download from a computer to the instrument. Technical Data |
| Device | SMD Boards, V-25 Microcomputer, 18-bit Analog-Digital Converter, 1 MBit Flash-EPROM, 1 MBit E² PROM |
| Power Supply | external Power Supply, Input voltage:
94 to 243 Vac, Input frequency: 50 to 60 Hz, Power output (max): 60
VA, Approvals: CSA, UL, TÜV-GS, CE-Mark |
| Optic block | 4-channel Optics with Halogen lamp and light intensity control, 16 different filters possible (coded), Range: 0-2,5 OD linear, up to 3 OD max., Resolution: 1 mOD |
| Keyboard | Foil keyboard, 4x8 matrix, with Test-, Functions- and numerical keys |
| Display | 4 lines x 20 Characters, Liquid Crystal Display (LCD) |
| Incubation block | 6 Cuvettes prewarming positions, 1
Measuring position, 1 Reagent position Ø 32mm, mixed, 2 Reagent positions Ø 22,5mm, 2 Reagent positions Ø 16 mm, Temperature: 37,0°C ± 0,2°C |
| Autopipette: (optional) |
3-stage Autopipette with electronical
start Volume: 50/100/200 µl |
| Printer: (optional) |
external, serial Thermal printer with print-out of all results, print-out of test parameters, print-out of system parameters, service print-out with all specific system information |
Standard package
|
Consumables and Accessory
|
| Reagents |
|
Important advantages
|